S.I. No. 160/2007 - Legal Metrology (European Conformity Assessment of Measuring Instruments) Regulations 2007
This is the act as published. Later amendments are not folded into the text below: each one is a separate act in this repository and an entry in this law's history.
2 amending acts · 2018-01-09, S.I. No. 2/2018 - Legal Metrology (European Conformity Assessment of Measuring Instrument…
— the definition of the metrological characteristics,
— the reproducibility of the metrological performances of produced measuring instruments when properly adjusted using appropriate intended means, and
— the integrity of the measuring instrument.
The technical documentation shall include insofar as relevant for assessment and identification of the type and/or measuring instrument—
- (a) a general description of the measuring instrument;
- (b) conceptual design and manufacturing drawings and plans of components, sub-assemblies, circuits, etc;
- (c) manufacturing procedures to ensure consistent production;
- (d) if applicable, a description of the electronic devices with drawings, diagrams, flowdiagrams of the logic and general software information explaining their characteristics and operation;
- (e) descriptions and explanations necessary for the understanding of paragraphs(b), (c) and(d), including the operation of the measuring instrument;
- (f) a list of the standards and/or normative documents referred to in paragraph (1) of Regulation 8, applied in full or in part;
- (g) descriptions of the solutions adopted to meet the essential requirements referred to in paragraph (1) of Regulation 6 where the standards and/or normative documents referred to in paragraph (1) of Regulation 8 have not been applied;
- (h) results of design calculations, examinations, etc;
- (i) the appropriate test results, where necessary, to demonstrate that the type and/or measuring instruments comply with:
— the requirements of these Regulations and the appropriate specific-instrument regulations under declared rated operating conditions and under specified environmental disturbances,
— the durability specifications for gas and water meters as well as meters for liquids other than water.
- (j) the EC type-examination certificates or EC design examination certificates in respect of measuring instruments containing parts identical to those in the design.
The manufacturer shall specify where seals and markings have been applied.
The manufacturer shall indicate the conditions for compatibility with interfaces and sub-assemblies, where relevant.
SCHEDULE V CRITERIA TO BE SATISFIED BY NOTIFIED BODIES
The following criteria apply for the designation of bodies in accordance with Regulation 12.
The body, its director and staff involved in conformity assessment tasks shall not be the designer, manufacturer, supplier, installer or user of the measuring instruments that they inspect, nor the authorised representative of any of them. In addition, they may not be directly involved in the design, manufacture, marketing or maintenance of the measuring instruments, nor represent the parties engaged in these activities. The preceding criterion does not, however, preclude in any way the possibility of exchanges of technical information between the manufacturer and the body for the purposes of conformity assessment.
The body, its director and staff involved in conformity assessment tasks shall be free from all pressures and inducements, in particular financial inducements, that might influence their judgement or the results of their conformity assessment, especially from persons or groups of persons with an interest in the results of the assessments.
The conformity assessment shall be carried out with the highest degree of professional integrity and requisite competence in the field of metrology. Should the body subcontract specific tasks, it shall first ensure that the subcontractor meets the requirements of these Regulations, and in particular this Schedule. The body shall keep the relevant documents assessing the subcontractor’s qualifications and the work carried out by him under these Regulations at the disposal of the Minister.
The body shall be capable of carrying out all the conformity assessment tasks for which it has been designated, whether those tasks are carried out by the body itself or on its behalf and under its responsibility. It shall have at its disposal the necessary staff and shall have access to the necessary facilities for carrying out in a proper manner the technical and administrative tasks entailed in conformity assessment.
The body’s staff shall have—
- (a) sound technical and vocational training, covering all conformity assessment tasks for which the body was designated;
- (b) satisfactory knowledge of the rules governing the tasks which it carries out, and adequate experience of such tasks;
- (c) the requisite ability to draw up the certificates, records and reports demonstrating that the tasks have been carried out.
The impartiality of the body, its director and staff shall be guaranteed. The remuneration of the body shall not depend on the results of the tasks it carries out. The remuneration of the body’s director and staff shall not depend on the number of tasks carried out or the results of such tasks.
The body shall take out civil liability insurance if its civil liability is not covered by national law.
The body’s director and staff shall be bound to observe professional secrecy with regard to all information obtained in the performance of their duties pursuant to these Regulations, except vis-à-vis the authority of the Minister.
SCHEDULE MI-001 WATER METERS FOR COLD WATER
The relevant requirements of Schedule III, the specific requirements of this Schedule and the conformity assessment procedures listed in this Schedule, apply to water meters intended for the measurement of volumes of clean, cold water in residential, commercial and light industrial use.
DEFINITIONS
Water Meter
An instrument designed to measure, memorise and display the volume at metering conditions of water passing through the measurement transducer.
Minimum Flowrate (Q1)
The lowest flowrate at which the water meter provides indications that satisfy the requirements concerning the maximum permissible errors (MPEs.)
Transitional Flowrate (Q2)
The transitional flowrate is the flowrate value occurring between the permanent and minimum flowrates, at which the flowrate range is divided into two zones, the ‘upper zone’ and the ‘lower zone’. Each zone has a characteristic MPE.
Permanent Flowrate (Q3)
The highest flowrate at which the water meter operates in a satisfactory manner under normal conditions of use, i.e. under steady or intermittent flow conditions.
Overload Flowrate (Q4)
The overload flowrate is the highest flowrate at which the meter operates in a satisfactory manner for a short period of time without deteriorating.
SPECIFIC REQUIREMENTS
Rated Operating Conditions
The manufacturer shall specify the rated operating conditions for the instrument, in particular;
The flowrate range of the water.
The values for the flowrate range shall fulfil the following conditions:
Q 3 /Q 1 ≥ 10
Q 2 /Q 1 = 1.6
Q 4 /Q 3 = 1.25
Until 30 April 2009, the ratio Q 2 /Q 1 may be: 1.5, 2.5, 4 or 6.3.
The temperature range of the water.
The values for the temperature range shall fulfil the following conditions:
1°C to at least 30°C, or
30°C to at least 90°C.
The meter may be designed to operate over both ranges.
The relative pressure range of the water, the range being 0.3 bar to at least 10 bar at Q3.
For the power supply: the nominal value of the AC voltage supply and/or the limits of DC supply.
MPE
The MPE, positive or negative, on volumes delivered at flowrates between the transitional flowrate (Q2) (included) and the overload flowrate (Q4) is:
2% for water having a temperature ≤ 30°C,
3% for water having a temperature > 30°C.
The MPE, positive or negative, on volumes delivered at flowrates between the minimum flowrate (Q1) and the transitional flowrate (Q2) (excluded) is 5% for water having any temperature.
Permissible Effect of Disturbances
- Electromagnetic immunity
1.1. The effect of an electromagnetic disturbance on a water meter shall be such that:
— the change in the measurement result is no greater than the critical change value as defined in 7.1.3, or
— the indication of the measurement result is such that it cannot be interpreted as a valid result, such as a momentary variation that cannot be interpreted, memorised or transmitted as a measuring result.
1.2. After undergoing an electromagnetic disturbance the water meter shall:
— recover to operate within MPE, and
— have all measurement functions safeguarded, and
— allow recovery of all measurement data present just before the disturbance.
1.3. The critical change value is the smaller of the two following values:
— the volume corresponding to half of the magnitude of the MPE in the upper zone on the measured volume;
— the volume corresponding to the MPE on the volume corresponding to one minute at flowrate Q3.
- Durability
After an appropriate test, taking into account the period of time estimated by the manufacturer, has been performed, the following criteria shall be satisfied:
2.1. The variation of the measurement result after the durability test, when compared with the initial measurement result, shall not exceed:
— 3% of the metered volume between Q1 included and Q2 excluded;
— 1.5% of the metered volume between Q2 included and Q4 included.
2.2. The error of indication for the volume metered after the durability test shall not exceed:
— ±6% of the metered volume between Q1 included and Q2 excluded;
— ±2.5% of the metered volume between Q2 included and Q4 included for water meters intended to meter water with a temperature between 0.1°C and 30°C,
— ±3.5% of the metered volume between Q2 included and Q4 included for water meters intended to meter water with a temperature between 30°C and 90°C.
Suitability
- The meter shall be able to be installed to operate in any position unless clearly marked otherwise.
- The manufacturer shall specify whether the meter is designed to measure reverse flow. In such a case, the reverse flow volume shall either be subtracted from the cumulated volume or shall be separately recorded. The same MPE shall apply to both forward and reverse flow.
Water meters not designed to measure reverse flow shall either prevent reverse flow or shall withstand an accidental reverse flow without any deterioration or change in metrological properties.
Units of Measurement
Metered volume shall be displayed in cubic metres.
CONFORMITY ASSESSMENT
The conformity assessment procedures referred to in Regulation 7 that the manufacturer can choose between are:
B followed by F, or
B followed by D,
or H1.
SCHEDULE MI-002 SPECIFIC REQUIREMENTS FOR GAS METERS AND VOLUME CONVERSION DEVICES
The relevant requirements of Schedule III, the specific requirements of this Schedule and the conformity assessment procedures listed in this Schedule, apply to gas meters and volume conversion devices defined below, intended for residential, commercial and light industrial use.
DEFINITIONS
Gas meter
An instrument designed to measure, memorise and display the quantity of fuel gas (volume or mass) that has passed it.
Conversion device
A device fitted to a gas meter that automatically converts the quantity measured at metering conditions into a quantity at base conditions.
Minimum flowrate (Qmin)
The lowest flowrate at which the gas meter provides indications that satisfy the requirements regarding maximum permissible error (MPE.)
Maximum flowrate (Qmax)
The highest flowrate at which the gas meter provides indications that satisfy the requirements regarding MPE.
Transitional flowrate (Qt)
The transitional flowrate is the flowrate occurring between the maximum and minimum flowrates at which the flowrate range is divided into two zones, the ‘upper zone’ and the ‘lower zone’. Each zone has a characteristic MPE.
Overload Flowrate (Qr)
The overload flowrate is the highest flowrate at which the meter operates for a short period of time without deteriorating.
Base conditions
The specified conditions to which the measured quantity of fluid is converted.
PART I SPECIFIC REQUIREMENTS — GAS METERS
Rated operating conditions
The manufacturer shall specify the rated operating conditions of the gas meter, taking into account:
- The flowrate range of the gas shall fulfil at least the following conditions:
| Class | Qmax/Qmin | Qmax/Qt | Qr/Qmax |
| 1.5 | ≥ 150 | ≥ 10 | 1.2 |
| 1.0 | ≥ 20 | ≥ 5 | 1.2 |
- The temperature range of the gas, with a minimum range of 40°C.
- The fuel/gas related conditions
The gas meter shall be designed for the range of gases and supply pressures of the country of destination. In particular the manufacturer shall indicate:
— the gas family or group;
— the maximum operating pressure.
- A minimum temperature range of 50°C for the climatic environment.
- The nominal value of the AC voltage supply and/or the limits of DC supply.
Maximum permissible error (MPEs)
- Gas meter indicating the volume at metering conditions or mass
| Class | 1.5 | 1.0 |
| Qmin ≤ Q < Qt | 3% | 2% |
| Qt ≤ Q ≤ Qmax | 1.5% | 1% |
Table 1
When the errors between Qt and Qmax all have the same sign, they shall all not exceed 1% for class 1.5 and 0.5% for Class 1.0.
- For a gas meter with temperature conversion, which only indicates the converted volume, the MPE of the meter is increased by 0.5% in a range of 30°C extending symmetrically around the temperature specified by the manufacturer that lies between 15°C and 25°C. Outside this range, an additional increase of 0.5% is permitted in each interval of 10°C.
Permissible effect of disturbances
- Electromagnetic immunity
1.1. The effect of an electromagnetic disturbance on a gas meter or volume conversion device shall be such that:
— the change in the measurement result is no greater than the critical change value as defined in 3.1.3, or
— the indication of the measurement result is such that it cannot be interpreted as a valid result, such as a momentary variation that cannot be interpreted, memorised or transmitted as a measuring result.
1.2. After undergoing a disturbance, the gas meter shall:
— recover to operate within MPE, and
— have all measurement functions safeguarded, and
— allow recovery of all measurement data present just before the disturbance.
1.3. The critical change value is the smaller of the two following values:
— the quantity corresponding to half of the magnitude of the MPE in the upper zone on the measured volume;
— the quantity corresponding to the MPE on the quantity corresponding to one minute at maximum flowrate.
- Effect of upstream-downstream flow disturbances
Under installation conditions specified by the manufacturer, the effect of the flow disturbances shall not exceed one third of the MPE.
Durability
After an appropriate test, taking into account the period of time estimated by the manufacturer, has been performed, the following criteria shall be satisfied:
- Class 1.5 meters
1.1. The variation of the measurement result after the durability test when compared with the initial measurement result for the flow rates in the range Qt to Qmax shall not exceed the measurement result by more than 2%.
1.2. The error of indication after the durability test shall not exceed twice the MPE in paragraph 2.
- Class 1.0 meters
2.1. The variation of the measurement result after the durability test when compared with the initial measurement result shall not exceed one-third of the MPE in paragraph 2.
2.2. The error of indication after the durability test shall not exceed the MPE in paragraph 2.
Suitability
- A gas meter powered from the mains (AC or DC) shall be provided with an emergency power supply device or other means to ensure, during a failure of the principal power source, that all measuring functions are safeguarded.
- A dedicated power source shall have a lifetime of at least five years. After 90% of its lifetime an appropriate warning shall be shown.
- An indicating device shall have a sufficient number of digits to ensure that the quantity passed during 8,000 hours at Qmax does not return the digits to their initial values.
- The gas meter shall be able to be installed to operate in any position declared by the manufacturer in its installation instruction.
- The gas meter shall have a test element, which shall enable tests to be carried out in a reasonable time.
- The gas meter shall respect the MPE in any flow direction or only in one flow direction clearly marked.
Units
Metered quantity shall be displayed in cubic metre, or in kilogram.
PART II SPECIFIC REQUIREMENTS — VOLUME CONVERSION DEVICES
A volume conversion device constitutes a sub-assembly of the second type according to the definition of sub-assembly given in Regulation 2.
For a volume conversion device, the essential requirements for the gas meter shall apply, if applicable. In addition, the following requirements shall apply:
Base conditions for converted quantities
The manufacturer shall specify the base conditions for converted quantities.
MPE
— 0.5% at ambient temperature 20°C ± 3°C, ambient humidity 60% ± 15%, nominal values for power supply;
— 0.7% for temperature conversion devices at rated operating conditions;
— 1% for other conversion devices at rated operating conditions.
Note: The error of the gas meter is not taken into account.
Suitability
- An electronic conversion device shall be capable of detecting when it is operating outside the operating range(s) stated by the manufacturer for parameters that are relevant for measurement accuracy. In such a case, the conversion device must stop integrating the converted quantity, and may totalise separately the converted quantity for the time it is operating outside the operating range(s).
- An electronic conversion device shall be capable to display all relevant data for the measurement without additional equipment.
PART III PUTTING INTO USE AND CONFORMITY ASSESSMENT
Putting into use
(a) Class 1.0 or Class 1.5 meters which have a Qmax/Qmin ratio greater than 150 may be used for measurement of gas in residential use.
- (b) Class 1.5 meters shall be used for measurement of commercial and/or light industrial use.
The distribution operator shall ensure the requirements under paragraphs 1.2 and 1.3 are determined in order that the meter installed is appropriate for the accurate measurement of consumption that is foreseen or foreseeable.
CONFORMITY ASSESSMENT
The conformity assessment procedures referred to in Regulation 7 that the manufacturer can choose between are—
B followed by F, or
B followed by D, or
H1.
SCHEDULE MI-003 ACTIVE ELECTRICAL ENERGY METERS
The relevant requirements of Schedule III, the specific requirements of this Schedule and the conformity assessment procedures listed in this Schedule, apply to active electrical energy meters intended for residential, commercial and light industrial use.
Note: Electrical energy meters may be used in combination with external instrument transformers, depending upon the measurement technique applied. However, this Schedule covers only electrical energy meters but not instrument transformers.
DEFINITIONS
An active electrical energy meter is a device which measures the active electrical energy consumed in a circuit.
I = the electrical current flowing through the meter;
In = the specified reference current for which the transformer operated meter has been designed;
Ist = the lowest declared value of I at which the meter registers active electrical energy at unity power factor (polyphase meters with balanced load);
Imin = the value of I above which the error lies within maximum permissible errors (MPEs) (polyphase meters with balanced load);
Itr = the value of I above which the error lies within the smallest MPE corresponding to the class index of the meter;
Imax = the maximum value of I for which the error lies within the MPEs;
U = the voltage of the electricity supplied to the meter;
Un = the specified reference voltage;
f = the frequency of the voltage supplied to the meter;
fn = the specified reference frequency;
PF = power factor = cos φ = the cosine of the phase difference φ between I and U.
SPECIFIC REQUIREMENTS
Accuracy
The manufacturer shall specify the class index of the meter. The class indices are defined as: Class A, B and C.
Rated operating conditions
The manufacturer shall specify the rated operating conditions of the meter; in particular:
The values of fn, Un, In, Ist, Imin, Itr and Imax that apply to the meter. For the current values specified, the meter shall satisfy the conditions given in Table 1;
| Class A | Class B | Class C | |
| For direct-connected meters | For direct-connected meters | For direct-connected meters | For direct-connected meters |
| Ist | ≤ 0.05 Itr | ≤ 0.04 Itr | ≤ 0.04 Itr |
| Imin | ≤ 0.5 Itr | ≤ 0.5 Itr | ≤ 0.3 Itr |
| Imax | ≥ 50 Itr | ≥ 50 Itr | ≥ 50 Itr |
| For transformer-operated meters | For transformer-operated meters | For transformer-operated meters | For transformer-operated meters |
| Ist | ≤ 0.06 Itr | ≤ 0.04 Itr | ≤ 0.02 Itr |
| Imin | ≤ 0.4 Itr | ≤ 0.2 Itr^*) | ≤ 0.2 Itr |
| In | = 20 Itr | = 20 Itr | = 20 Itr |
| Imax | ≥ 1.2 In | ≥ 1.2 In | ≥ 1.2 In |
^*) For Class B electromechanical meters Imin ≤ 0.4 Itr shall apply.
Table 1
The voltage, frequency and power factor ranges within which the meter shall satisfy the MPE requirements are specified in Table 2. These ranges shall recognise the typical characteristics of electricity supplied by public distribution systems.
The voltage and frequency ranges shall be at least:
9 Un ≤ U ≤ 1.1 Un
98 fn ≤ f ≤ 1.02 fn
power factor range at least from cosφ = 0.5 inductive to cosφ = 0.8 capacitive.
MPEs
The effects of the various measurands and influence quantities (a, b, c,...) are evaluated separately, all other measurands and influence quantities being kept relatively constant at their reference values. The error of measurement, that shall not exceed the MPE stated in Table 2, is calculated as:
Error of measurement = [image omitted]
When the meter is operating under varying-load current, the percentage errors shall not exceed the limits given in Table 2.
For electromechanical polyphase meters the current range for single-phase load is limited to 5Itr ≤ I ≤ Imax
Table 2 — MPEs in percent at rated operating conditions and defined load current levels and operating temperature
When a meter operates in different temperature ranges the relevant MPE values shall apply.
Permissible effect of disturbances
- General
As electrical energy meters are directly connected to the mains supply and as mains current is also one of the measurands, a special electromagnetic environment is used for electricity meters.
The meter shall comply with the electromagnetic environment E2 and the additional requirements in 4.2 and 4.3.
The electromagnetic environment and permissible effects reflect the situation that there are disturbances of long duration which shall not affect the accuracy beyond the critical change values and transient disturbances, which may cause a temporary degradation or loss of function or performance but from which the meter shall recover and shall not affect the accuracy beyond the critical change values.
When there is a foreseeable high risk due to lightning or where overhead supply networks are predominant, the metrological characteristics of the meter shall be protected.
- Effect of disturbances of long duration
| Disturbance | Critical change values in percent for meters of class | Critical change values in percent for meters of class | Critical change values in percent for meters of class |
| A | B | C | |
| Reversed phase sequence | 1. 5 | 1. 5 | 0. 3 |
| Voltage unbalance (only applicable to polyphase meters) | 4 | 2 | 1 |
| Harmonic contents in the current circuits * | 1 | 0. 8 | 0. 5 |
| DC and harmonics in the current circuit * | 6 | 3 | 1. 5 |
| Fast transient bursts | 6 | 4 | 2 |
| Magnetic fields; HF (radiated RF) electromagnetic field; Conducted disturbances introduced by radio-frequency fields; and Oscillatory waves immunity | 3 | 2 | 1 |
- In the case of electromechanical electricity meters, no critical change values are defined for harmonic contents in the current circuits and for DC and harmonics in the current circuit.
Table 3
Critical change values for disturbances of long duration
- Permissible effect of transient electromagnetic phenomena
3.1. The effect of an electromagnetic disturbance on an electrical energy meter shall be such that during and immediately after a disturbance
— any output intended for testing the accuracy of the meter does not produce pulses or signals corresponding to an energy of more than the critical change value and in reasonable time after the disturbance the meter shall
— recover to operate within the MPE limits, and
— have all measurement functions safeguarded, and
— allow recovery of all measurement data present prior to the disturbance, and
— not indicate a change in the registered energy of more than the critical change value.
The critical change value in kWh is m . Un . Imax . 10-6
(m being the number of measuring elements of the meter, Un in Volts and Imax in Amps).
3.2. For overcurrent the critical change value is 1.5%.
Suitability
- Below the rated operating voltage the positive error of the meter shall not exceed 10%.
- The display of the total energy shall have a sufficient number of digits to ensure that when the meter is operated for 4,000 hours at full load (I = Imax, U = Un and PF = 1) the indication does not return to its initial value and shall not be able to be reset during use.
- In the event of loss of electricity in the circuit, the amounts of electrical energy measured shall remain available for reading during a period of at least 4 months.
- Running with no load
When the voltage is applied with no current flowing in the current circuit (current circuit shall be open circuit), the meter shall not register energy at any voltage between 0.8 ^. Un and 1.1 Un.
- Starting
The meter shall start and continue to register at Un PF = 1 (polyphase meter with balanced loads) and a current which is equal to Ist.
Units
The electrical energy measured shall be displayed in kilowatt-hours or in megawatt-hours.
Putting into use
- (a) For residential use, Class A meters may be used. For specified purposes, Class B meters may be required.
- (b) For commercial and / or light industrial use, Class B meters may be used. For specified purposes, Class C meters may be required.
- (c) The current range shall be determined by the distributor or the person legally designated for installing the meter, so that the meter is appropriate for the accurate measurement of consumption that is foreseen or foreseeable.
CONFORMITY ASSESSMENT
The conformity assessment procedures referred to in Regulation 7 that the manufacturer can choose between are—
B followed by F, or
B followed by D, or
H1.
SCHEDULE MI-005 MEASURING SYSTEMS FOR THE CONTINUOUS AND DYNAMIC MEASUREMENT OF QUANTITIES OF LIQUIDS OTHER THAN WATER
The relevant essential requirements of Schedule III, the specific requirements of this Schedule and the conformity assessment procedures listed in this Schedule, apply to measuring systems intended for the continuous and dynamic measurement of quantities (volumes or masses) of liquids other than water as given in Table 1. If appropriate, the terms ‘volume, and L’ in this Schedule can be read as ‘mass and kg’.
DEFINITIONS
Meter
An instrument designed to measure continuously, memorise and display the quantity at metering conditions of liquid flowing through the measurement transducer in a closed, fully charged conduit.
Calculator
A part of a meter that receives the output signals from the measurement transducer(s) and possibly, from associated measuring instruments and displays the measurement results.
Associated measuring instrument
An instrument connected to the calculator for measuring certain quantities which are characteristic of the liquid, with a view to make a correction and/or conversion.
Conversion Device
A part of the calculator which by taking account of the characteristics of the liquid (temperature, density, etc.) measured using associated measuring instruments, or stored in a memory, automatically converts:
— the volume of the liquid measured at metering conditions into a volume at base conditions and/or into mass, or
— the mass of the liquid measured at metering conditions into a volume at metering conditions and/or into a volume at base conditions
Note: A conversion device includes the relevant associated measuring instruments.
Base conditions
The specified conditions to which the measured quantity of liquid at metering conditions is converted.
Measuring System
A system that comprises the meter itself and all devices required to ensure correct measurement or intended to facilitate the measuring operations.
Fuel dispenser
A measuring system intended for the refuelling of motor vehicles, small boats and small aircraft.
Self-service arrangement
An arrangement that allows the customer to use a measuring system for the purpose of obtaining liquid for his own use.
Self-service device
A specific device that is part of a self-service arrangement and which allows one of more measuring systems to perform in this self-service arrangement.
Minimum measured quantity (MMQ)
The smallest quantity of liquid for which the measurement is metrologically acceptable for the measuring system.
Direct indication
The indication, either volume or mass, corresponding to the measure and that the meter is physically capable of measuring.
Note: The direct indication may be converted into another quantity using a conversion device.
Interruptible/non interruptible
A measuring system is considered as interruptible/non interruptible when the liquid flow can/cannot be stopped easily and rapidly.
Flowrate range
The range between the minimum flowrate (Qmin) and maximum flowrate (Qmax).
SPECIFIC REQUIREMENTS
Rated operating conditions
The manufacturer shall specify the rated operating conditions for the instrument, in particular;
- The flowrate range:
The flowrate range is subject to the following conditions:
- (i) the flowrate range of a measuring system shall be within the flowrate range of each of its elements, in particular the meter.
- (ii) meter and measuring system:
| Specific measuring system | Characteristic of liquid | Minimum ratio of Qmax: Qmin |
| Fuel dispensers | Not Liquefied gases | 10 : 1 |
| —Measuring systems on road tankers for liquid fuel | Liquid fuel or milk | 4 : 1 |
| —Measuring systems for milk |
Table 1
- The properties of the liquid to be measured by the instrument by specifying the name or type of the liquid or its relevant characteristics, for example:
— Temperature range;
— Pressure range;
— Density range;
— Viscosity range.
- The nominal value of the AC voltage supply and/or limits of the DC voltage supply.
- The base conditions for converted values.
Note: Paragraph 1.4 is without prejudice to the Member States’ obligations to require use of a temperature of either 15°C in accordance with Article 3(1) of Council Directive 92/81/EEC of 19 October 1992 on the harmonisation of the structures of excise duties on mineral oil 2 or, for heavy fuel oils, LPG and methane, another temperature pursuant to Article 3(2) of that Directive.
1 OJ L 316, 31.10.1992, p. 12. Directive repealed by Directive 2003/96/EC (OJ L 283, 31.10.2003, p. 51).
Accuracy classification and maximum permissible errors (MPEs)
- For quantities equal to or greater than 2 litres the MPE on indications is:
| Accuracy Class 0.5 | |
| Measuring systems (A) | 0. 5% |
| Meters (B) | 0. 3% |
Table 2
- For quantities less than two litres the MPE on indications is:
| Measured volume V | MPE |
| V < 0.1 L | 4 x value in Table 2, applied to 0.1 L |
| 0. 1 L ≤ V < 0.2 L | 4 x value in Table 2 |
| 0. 2 L ≤ V < 0.4 L | 2 x value in Table 2, applied to 0.4 L |
| 0. 4 L ≤ V < 1 L | 2 x value in Table 2 |
| 1 L ≤ V < 2 L | Value in Table 2, applied to 2 L |
Table 3
- However, no matter what the measured quantity may be, the magnitude of the MPE is given by the greater of the following two values:
— the absolute value of the MPE given in Table 2 or Table 3,
— the absolute value of the MPE for the minimum measured quantity (Emin).
4.1. For minimum measured quantities greater than or equal to 2 litres the following conditions apply:
Condition 1
Emin shall fulfil the condition: Emin ≥ 2 R, where R is the smallest scale interval of the indication device.
Condition 2
Emin is given by the formula: Emin = (2MMQ) x (A/100), where:
— MMQ is the minimum measured quantity,
— A is the numerical value specified in line A of Table 2.
4.2. For minimum measured quantities of less than two litres, the above-mentioned condition 1 applies and Emin is twice the value specified in Table 3, and related to line A of Table 2.
- Converted indication
In the case of a converted indication the MPEs are as in line A of Table 2.
- Conversion devices
MPEs on converted indications due to a conversion device are equal to ±(A — B), A and B being the values specified in Table 2.
Parts of conversion devices that can be tested separately
- (a) Calculator
MPEs on quantities of liquid indications applicable to calculation, positive or negative, are equal to one-tenth of the MPEs as defined in line A of Table 2.
- (b) Associated measuring instruments
Associated measuring instruments shall have an accuracy at least as good as the values in Table 4:
| MPE on Measurements for measuring system of 0.5% | MPE on Measurements for measuring system of 0.5% |
| Temperature | ± 0.5°C |
| Pressure | Less than 1 MPa: ± 50 kPa |
| From 1 to 4 MPa: ± 5% | |
| Over 4 MPa: ± 200 kPa | |
| Density | ± 1 kg/m^3 |
Table 4
These values apply to the indication of the characteristic quantities of the liquid displayed by the conversion device.
- (c) Accuracy for calculating function
The MPE for the calculation of each characteristic quantity of the liquid, positive or negative, is equal to two fifths of the value fixed in (b).
- The requirement (a) in paragraph 2.6 applies to any calculation, not only conversion.
Maximum permissible effect of disturbances
- The effect of an electromagnetic disturbance on a measuring system shall be one of the following:
— the change in the measurement result is not greater than the critical change value as defined in paragraph 3.2, or
— the indication of the measurement result shows a momentary variation that cannot be interpreted, memorised or transmitted as a measuring result. Furthermore, in the case of an interruptible system, this can also mean the impossibility to perform any measurement, or
— the change in the measurement result is greater than the critical change value, in which case the measuring system shall permit the retrieval of the measuring result just before the critical change value occurred and cut off the flow.
- The critical change value is the greater of MPE/5 for a particular measured quantity or Emin.
Durability
After an appropriate test, taking into account the period of time estimated by the manufacturer, has been performed, the following criterion shall be satisfied:
The variation of the measurement result after the durability test, when compared with the initial measurement result, shall not exceed the value for meters specified in line B of table 2.
Suitability
- For any measured quantity relating to the same measurement, the indications provided by various devices shall not deviate one from another by more than one scale interval where devices have the same scale interval. In the case where the devices have different scale intervals, the deviation shall not be more than that of the greatest scale interval. However, in the case of a self-service arrangement the scale intervals of the main indicating device on the measuring system and the scale intervals of the self-service device shall be the same and results of measurement shall not deviate one from another.
- It shall not be possible to divert the measured quantity in normal conditions of use unless it is readily apparent.
- Any percentage of air or gas not easily detectable in the liquid shall not lead to a variation of error greater than:
— 0.5% for liquids other than potable liquids and for liquids of a viscosity not exceeding 1 mPa.s, or
— 1% for potable liquids and for liquids of a viscosity exceeding 1 mPa.s.
However, the allowed variation shall never be smaller than 1% of MMQ. This value applies in the case of air or gas pockets.
- Instruments for direct sales
4.1. A measuring system for direct sales shall be provided with means for resetting the display to zero.
It shall not be possible to divert the measured quantity.
4.2. The display of the quantity on which the transaction is based shall be permanent until all parties in the transaction have accepted the measurement result.
4.3. Measuring systems for direct sales shall be interruptible.
4.4. Any percentage of air or gas in the liquid shall not lead to a variation of error greater than the values specified in paragraph 5.3.
- Fuel Dispensers
5.1. Displays on fuel dispensers shall not be capable of being reset to zero during a measurement.
5.2. The start of a new measurement shall be inhibited until the display has been reset to zero.
5.3. Where a measuring system is fitted with a price display, the difference between the indicated price and the price calculated from the unit price and the indicated quantity shall not exceed the price corresponding to Emin.
However this difference need not be less than the smallest monetary value.
Power supply failure
A measuring system shall either be provided with an emergency power supply device that will safeguard all measuring functions during the failure of the main power supply device or be equipped with means to save and display the data present in order to permit the conclusion of the transaction in progress and with means to stop the flow at the moment of the failure of the main power supply device.
7. Putting into use
| Accuracy Class | Types of Measuring system |
| 0. 5 | — fuel dispensers (not for liquefied gases), |
| — measuring systems on road tankers for liquid fuel | |
| — measuring systems for milk |
Table 5
Note: However, the manufacturer may specify a better accuracy for a certain type of measuring system.
Units of measurement
The metered quantity shall be displayed in millilitres, cubic centimetres, litres, cubic metres, grams, kilograms or tonnes.
CONFORMITY ASSESSMENT
The conformity assessment procedures referred to in Regulation 7 that the manufacturer can choose between are—
B followed by F, or
B followed by D, or
H1, or
G.
SCHEDULE MI-006 AUTOMATIC WEIGHING INSTRUMENTS
The relevant essential requirements of Schedule III, the specific requirements of this Schedule and the conformity assessment procedures listed in Chapter I of this Schedule, apply to automatic weighing instruments defined below, intended to determine the mass of a body by using the action of gravity on that body.
DEFINITIONS
Automatic weighing instrument
An instrument that determines the mass of a product without the intervention of an operator and follows a predetermined programme of automatic processes characteristic of the instrument.
Automatic catchweigher
An automatic weighing instrument that determines the mass of pre-assembled discrete loads (for example pre-packages) or single loads of loose material.
Automatic checkweigher
An automatic catchweigher that subdivides articles of different mass into two or more subgroups according to the value of the difference of their mass and a nominal set-point.
Weight labeller
An automatic catchweigher that labels individual articles with the weight value.
Weight/price labeller
An automatic catchweigher that labels individual articles with the weight value, and price information.
Automatic gravimetric filling instrument
An automatic weighing instrument that fills containers with a predetermined and virtually constant mass of product from bulk.
Discontinuous totaliser (totalising hopper weigher)
An automatic weighing instrument that determines the mass of a bulk product by dividing it into discrete loads. The mass of each discrete load is determined in sequence and summed. Each discrete load is then delivered to bulk.
Continuous totaliser
An automatic weighing instrument that continuously determines the mass of a bulk product on a conveyor belt, without systematic subdivision of the product and without interrupting the movement of the conveyor belt.
SPECIFIC REQUIREMENTS
CHAPTER I Requirements common to all types of automatic weighing instruments
Rated Operating Conditions
The manufacturer shall specify the rated operating conditions for the instrument as follows:
- For the measurand:
The measuring range in terms of its maximum and minimum capacity.
- For the electrical supply influence quantities:
| In case of AC voltage supply: | the nominal AC voltage supply, or the AC voltage limits. |
| In case of DC voltage supply: | the nominal and minimum DC voltage supply, or the DC voltage limits. |
- For the mechanical and climatic influence quantities:
The minimum temperature range is 30°C unless specified otherwise in the following chapters of this Schedule.
The mechanical environment classes according to Schedule III, paragraph 1.3.2 are not applicable. For instruments which are used under special mechanical strain, e.g. instruments incorporated into vehicles, the manufacturer shall define the mechanical conditions of use.
- For other influence quantities (if applicable)—
— The rate(s) of operation.
— The characteristics of the product(s) to be weighed.
Permissible effect of disturbances — Electromagnetic environment
The required performance and the critical change value are given in the relevant Chapter of this Schedule for each type of instrument.
Suitability
- Means shall be provided to limit the effects of tilt, loading and rate of operation such that maximum permissible errors (MPEs) are not exceeded in normal operation.
- Adequate material handling facilities shall be provided to enable the instrument to respect the MPEs during normal operation.
- Any operator control interface shall be clear and effective.
- The integrity of the display (where present) shall be verifiable by the operator.
- Adequate zero setting capability shall be provided to enable the instrument to respect the MPEs during normal operation.
- Any result outside the measurement range shall be identified as such, where a printout is possible.
Conformity assessment
The conformity assessment procedures referred to in Regulation 7 that the manufacturer can choose between are:
For mechanical systems:
B followed by D, or
B followed by E, or
B followed by F, or
D1, or
F1, or
G, or
H1.
For electromechanical instruments:
B followed by D, or
B followed by E, or
B followed by F, or
G, or
H1.
For electronic systems or systems containing software:
B followed by D, or
B followed by F, or
G, or
H1.
CHAPTER II Automatic Catchweigher
Accuracy Classes
- Instruments are divided into primary categories designated by:
X or Y as specified by the manufacturer.
- These primary categories are further divided into four accuracy classes:
XI, XII, XIII & XIV and
Y(I), Y(II), Y(a) & Y(b)
which shall be specified by the manufacturer.
Category X Instruments
- Category X applies to instruments used to check pre-packages made up in accordance with the requirements of Council Directive 75/106/EEC of 19 December 1974 on the approximation of the laws of the Member States relating to the making-up by volume of certain pre-packaged liquids 3 and of Council Directive 76/211/EEC of 20 January 1976 on the approximation of the laws of the Member States relating to the making-up by weight or by volume of certain pre-packaged products 2 applicable to pre-packages.
1 OJ L 42, 15.2.1975, p. 1. Directive as last amended by Directive 89/676/EEC (OJ L 398, 30.12.1989, p. 18).
2 OJ L 46, 21.2.1976, p. 1. Directive as last amended by the EEA Agreement.
- The accuracy classes are supplemented by a factor (x) that quantifies the maximum permissible standard deviation as specified in paragraph 4.2.
The manufacturer shall specify the factor (x), where (x) shall be ≤ 2 and in the form 1 x 10k, 2 x 10k or 5 x 10k, where k is a negative whole number or zero.
Category Y Instruments
Category Y applies to all other automatic catchweighers.
MPE
- Mean error Category X / MPE Category Y instruments
Table 1
- Standard deviation
Maximum permissible value for the standard deviation of a class X (x) instrument is the result of the multiplication of the factor (x) by the value in Table 2 below.
| Net Load (m) | Maximum permissible standard deviation for class X(1) |
| m ≤ 50 g | 0.48% |
| 50 g < m ≤ 100 g | 0.24 g |
| 100 g < m ≤ 200 g | 0.24% |
| 200 g < m ≤ 300 g | 0.48 g |
| 300 g < m ≤ 500 g | 0.16% |
| 500 g < m ≤ 1,000 g | 0.8 g |
| 1,000 g < m ≤ 10,000 g | 0.08% |
| 10,000 g < m ≤ 15,000 g | 8 g |
| 15,000 g < m | 0. 053% |
| For class XI and XII (x) shall be less than 1. | For class XI and XII (x) shall be less than 1. |
| For class XIII (x) shall be not greater than 1. | For class XIII (x) shall be not greater than 1. |
| For class XIV (x) shall be greater than 1. | For class XIV (x) shall be greater than 1. |
Table 2
- Verification scale interval — single interval instruments
| Accuracy classes | Accuracy classes | Verification scale interval | Number of verification scale intervals n = Max/e | Number of verification scale intervals n = Max/e |
| Minimum | Maximum | |||
| XI | Y(I) | 0.001 g ≤ e | 50,000 | — |
| XII | Y(II) | 0.001 g ≤ e ≤ 0.05 g | 100 | 100,000 |
| 0.1 g ≤ e | 5,000 | 100,000 | ||
| XIII | Y(a) | 0.1 g ≤ e ≤ 2 g | 100 | 10,000 |
| 5 g ≤ e | 500 | 10,000 | ||
| XIV | Y(b) | 5 g ≤ e | 100 | 1,000 |
Table 3
- Verification scale interval — multi - interval instruments
| Accuracy classes | Accuracy classes | Verification scale interval | Number of verification scale intervals n = Max/e | Number of verification scale intervals n = Max/e |
| Minimum value^(1) n =Maxi/e(i+1) | Maximum valuen = Maxi/ei | |||
| XI | Y(I) | 0.001 g ≤ ei | 50,000 | — |
| XII | Y(II) | 0.001 g ≤ ei ≤ 0.05 g | 5,000 | 100,000 |
| 0.1 g ≤ ei | 5,000 | 100,000 | ||
| XIII | Y(a) | 0.1 g ≤ ei | 500 | 10,000 |
| XIV | Y(b) | 5 g ≤ ei | 50 | 1,000 |
Where: i = 1, 2,... r
i = partial weighing range
r = total number of partial ranges
- (1) For i = r the corresponding column of Table 3 applies with e replaced by er.
Table 4
Measurement Range
In specifying the measurement range for class Y instruments the manufacturer shall take account that the minimum capacity shall not be less than:
| class Y(I): | 100 e |
| class Y(II): | 20 e for 0.001 g ≤ e ≤ 0.05 g |
| and 50 e for 0.1 g ≤ e | |
| class Y(a): | 20 e |
| class Y(b): | 10 e |
| Scales used for grading, | |
| e.g. postal scales and waste weighers: | 5 e |
Dynamic Setting
- The dynamic setting facility shall operate within a load range specified by the manufacturer.
- When fitted, a dynamic setting facility that compensates for the dynamic effects of the load in motion shall be inhibited from operating outside the load range, and shall be capable of being secured.
Performance Under Influence Factors And Electromagnetic Disturbances
- The MPEs due to influence factors are:
1.1. For category X instruments:
— For automatic operation; as specified in Tables 1 and 2,
— For static weighing in non-automatic operation; as specified in Table 1.
1.2. For category Y instruments
— For each load in automatic operation; as specified in Table 1,
— For static weighing in non-automatic operation; as specified for category X in Table 1.
- The critical change value due to a disturbance is one verification scale interval.
- Temperature range:
— For class XI and Y(I) the minimum range is 5° C,
— For class XII and Y(II) the minimum range is 15° C.
CHAPTER III Automatic Gravimetric Filling Instruments
Accuracy classes
- The manufacturer shall specify both the reference accuracy class Ref(x) and the operational accuracy class(es) X(x).
- An instrument type is designated a reference accuracy class, Ref(x), corresponding to the best possible accuracy for instruments of the type. After installation, individual instruments are designated for one or more operational accuracy classes, X(x), having taken account of the specific products to be weighed. The class designation factor (x) shall be ≤ 2, and in the form 1 x 10k, 2 x 10k or 5 x 10k where k is a negative whole number or zero.
- The reference accuracy class, Ref(x) is applicable for static loads.
- For the operational accuracy class X(x), X is a regime relating accuracy to load weight and (x) is a multiplier for the limits of error specified for class X(1) in 2.2.
MPE
- Static weighing error
1.1. For static loads under rated operating conditions, the MPE for reference accuracy class Ref(x), shall be 0.312 of the maximum permissible deviation of each fill from the average; as specified in Table 5; multiplied by the class designation factor (x).
1.2. For instruments where the fill may be made up from more than one load (e.g. cumulative or selective combination weighers) the MPE for static loads shall be the accuracy required for the fill as specified in 2.2 (i.e. not the sum of the maximum permissible deviation for the individual loads).
- Deviation from average fill
| Value of the mass, m (g), of the fills | Maximum permissible deviation of each fill from the average for class X(1) |
| m ≤ 50 | 7.2% |
| 50 < m ≤ 100 | 3.6 g |
| 100 < m ≤ 200 | 3.6% |
| 200 < m ≤ 300 | 7.2 g |
| 300 < m ≤ 500 | 2.4% |
| 500 < m ≤ 1,000 | 12 g |
| 1,000 < m ≤ 10,000 | 1.2% |
| 10,000 < m ≤ 15,000 | 120 g |
| 15,000 < m | 0.8% |
| Note: The calculated deviation of each fill from the average may be adjusted to take account for the effect of material particle size. | Note: The calculated deviation of each fill from the average may be adjusted to take account for the effect of material particle size. |
Table 5
- Error relative to pre-set value (setting error)
For instruments where it is possible to pre-set a fill weight; the maximum difference between the pre-set value and the average mass of the fills shall not exceed 0.312 of the maximum permissible deviation of each fill from the average, as specified in Table 5.
Performance Under Influence Factor And Electromagnetic Disturbance
- The MPE due to influence factors shall be as specified in paragraph 2.1.
- The critical change value due to a disturbance is a change of the static weight indication equal to the MPE as specified in paragraph 2.1 calculated for the rated minimum fill, or a change that would give equivalent effect on the fill in the case of instruments where the fill consists of multiple loads. The calculated critical change value shall be rounded to the next higher scale interval (d).
- The manufacturer shall specify the value of the rated minimum fill.
CHAPTER IV Discontinuous Totalisers
Accuracy Classes
Instruments are divided into four accuracy classes as follows: 0.2, 0.5, 1 and 2.
MPEs
| Accuracy class | MPE of totalised load |
| 0.2 | ± 0.10% |
| 0.5 | ± 0.25% |
| 1 | ± 0.50% |
| 2 | ± 1.00% |
Table 6
Totalisation scale interval
The totalisation scale interval (dt) shall be in the range—
0.01% Max ≤ dt ≤ 0.2% Max
Minimum Totalised Load (Σmin)
The minimum totalised load (Σmin) shall be not less than the load at which the MPE is equal to the totalisation scale interval (dt) and not less than the minimum load as specified by the manufacturer.
Zero Setting
Instruments that do not tare weigh after each discharge shall have a zero setting device. Automatic operation shall be inhibited if zero indication varies by:
— 1 dt on instruments with automatic zero setting device;
— 0.5 dt on instruments with a semi-automatic, or non-automatic, zero setting device.
Operator Interface
Operator adjustments and reset function shall be inhibited during automatic operation.
Printout
On instruments equipped with a printing device, the reset of the total shall be inhibited until the total is printed.
The printout of the total shall occur if automatic operation is interrupted.
Performance under influence factors and electromagnetic disturbances
- The MPEs due to influence factors shall be as specified in Table 7 below.
| Load (m) in totalisation scale intervals (dt) | MPE |
| 0 < m ≤ 500 | ± 0.5 dt |
| 500 < m ≤ 2,000 | ± 1.0 dt |
| 2,000 < m ≤ 10,000 | ± 1.5 dt |
Table 7
- The critical change value due to a disturbance is one totalisation scale interval for any weight indication and any stored total.
CHAPTER V Continuous Totalisers
Accuracy classes
Instruments are divided into three accuracy classes as follows: 0.5, 1 and 2.
Measurement Range
- The manufacturer shall specify the measurement range, the ratio between the minimum net load on the weighing unit and the maximum capacity, and the minimum totalised load.
- The minimum totalised load Σmin shall not be less than
800 d for class 0.5,
400 d for class 1,
200 d for class 2.
Where d is the totalisation scale interval of the general totalisation device.
MPE
| Accuracy class | MPE for totalised load |
| 0.5 | ± 0.25% |
| 1 | ± 0.5% |
| 2 | ± 1.0% |
Table 8
Speed of the belt
The speed of the belt shall be specified by the manufacturer. For single-speed beltweighers, and variable-speed beltweighers having a manual speed setting control, the speed shall not vary by more than 5% of the nominal value. The product shall not have a different speed than the speed of the belt.
General Totalisation Device
It shall not be possible to reset the general totalisation device to zero.
Performance under influence factors and electromagnetic disturbances
- The MPE due to influence factor, for a load not less than the Σmin, shall be 0.7 times the appropriate value specified in Table 8, rounded to the nearest totalisation scale interval (d).
- The critical change value due to a disturbance shall be 0.7 times the appropriate value specified in Table 8, for a load equal to Σmin, for the designated class of the beltweigher; rounded up to the next higher totalisation scale interval (d).
SCHEDULE MI-007 TAXIMETERS
The relevant requirements of Schedule III, the specific requirements of this Schedule and the conformity assessment procedures listed in this Schedule apply to taximeters.
DEFINITIONS
Taximeter
A device that works together with a signal generator 5 to make a measuring instrument.
1 The distance signal generator is outside the scope of these Regulations.
This device measures duration, calculates distance on the basis of a signal delivered by the distance signal generator.
Additionally, it calculates and displays the fare to be paid for a trip on the basis of the calculated distance and/or the measured duration of the trip.
Fare
The total amount of money due for a trip based on a fixed initial hire fee and/or the length and/or the duration of the trip. The fare does not include a supplement charged for extra services.
Cross-over speed
The speed value found by division of a time tariff value by a distance tariff value.
Normal calculation mode S (single application of tariff)
Fare calculation based on application of the time tariff below the cross-over speed and application of the distance tariff above the cross-over speed.
Normal calculation mode D (double application of tariff)
Fare calculation based on simultaneous application of time tariff and distance tariff over the whole trip.
Operating position
The different modes in which a taximeter fulfils the different parts of its functioning. The operating positions are distinguished by the following indications:
‘For Hire’: The operating position in which the fare calculation is disabled
‘Hired’: The operating position in which the fare calculation takes place on the basis of a possible initial charge and a tariff for distance travelled and/or time of the trip
‘Stopped’: The operating position in which the fare due for the trip is indicated and at least the fare calculation based on time is disabled.
DESIGN REQUIREMENTS
The taximeter shall be designed to calculate the distance and to measure the duration of a trip.
The taximeter shall be designed to calculate and display the fare, incrementing in steps equal to the resolution fixed by the Member State in the operation position ‘Hired’. The taximeter shall also be designed to display the final value for the trip in the operating position ‘Stopped’.
A taximeter shall be able to apply the normal calculation modes S and D. It shall be possible to choose between these calculation modes by a secured setting.
A taximeter shall be able to supply the following data through an appropriate secured interface(s):
— operation position: ‘For Hire’, ‘Hired’ or ‘Stopped’;
— totaliser data according to paragraph 15.1;
— general information: constant of the distance signal generator, date of securing, taxi identifier, real time, identification of the tariff;
— fare information for a trip: total charged, fare, calculation of the fare, supplement charge, date, start time, finish time, distance travelled;
— tariff(s) information: parameters of tariff(s).
National legislation may require certain devices to be connected to the interface(s) of a taximeter. Where such a device is required; it shall be possible, by secured setting, to inhibit automatically the operation of the taximeter for reasons of the non-presence or improper functioning of the required device.
If relevant, it shall be possible to adjust a taximeter for the constant of the distance signal generator to which it is to be connected and to secure the adjustment.
RATED OPERATING CONDITIONS
- The mechanical environment class that applies is M3.
- The manufacturer shall specify the rated operating conditions for the instrument, in particular:
— a minimum temperature range of 80°C for the climatic environment;
— the limits of the DC power supply for which the instrument has been designed.
MAXIMUM PERMISSIBLE ERRORS (MPEs)
The MPE, excluding any errors due to application of the taximeter in a taxi, are:
— For the time elapsed:± 0.1%
minimum value of mpe: 0.2s;
— For the distance travelled:± 0.2%
minimum value of mpe: 4 m;
— For the calculation of the fare:± 0.1%
minimum, including rounding: corresponding to the least significant digit of the fare indication.
PERMISSIBLE EFFECT OF DISTURBANCES
Electromagnetic immunity
- The electromagnetic class that applies is E3.
- The MPE laid down in paragraph 7 shall also be respected in the presence of an electromagnetic disturbance.
POWER SUPPLY FAILURE
In case of a reduction of the voltage supply to a value below the lower operating limit as specified by the manufacturer, the taximeter shall:
— continue to work correctly or resume its correct functioning without loss of data available before the voltage drop if the voltage drop is temporary, i.e. due to restarting the engine;
— abort an existing measurement and return to the position ‘For Hire’ if the voltage drop is for a longer period.
OTHER REQUIREMENTS
The conditions for the compatibility between the taximeter and the distance signal generator shall be specified by the manufacturer of the taximeter.
If there is a supplement charge for an extra service, entered by the driver on manual command, this shall be excluded from the fare displayed. However, in that case a taximeter may display temporarily the value of the fare including the supplementary charge.
If the fare is calculated according to calculation mode D a taximeter may have an additional display mode in which only the total distance and duration of the trip are displayed in real time.
All values displayed for the passenger shall be suitably identified. These values as well as their identification shall be clearly readable under daylight and night conditions.
- If the fare to be paid or the measures to be taken against fraudulent use can be affected by the choice of functionality from a pre-programmed setting or by free data setting, it shall be possible to secure the instrument settings and data entered.
- The securing possibilities available in a taximeter shall be such that separate securing of the settings is possible.
- The provisions in paragraph 8.3 of Schedule III apply also to the tariffs.
- A taximeter shall be fitted with non-resettable totalisers for all of the following values:
— The total distance travelled by the taxi;
— The total distance travelled when hired;
— The total number of hirings;
— The total amount of money charged as supplements;
— The total amount of money charged as fare.
The totalised values shall include the values saved according to paragraph 9 under conditions of loss of power supply.
- If disconnected from power, a taximeter shall allow the totalised values to be stored for one year for the purpose of reading out the values from the taximeter to another medium.
- Adequate measures shall be taken to prevent the display of totalised values from being used to deceive passengers.
Automatic change of tariffs is allowed due to the:
— distance of the trip;
— duration of the trip;
— time of the day;
— date;
— day of the week.
If properties of the taxi are important for the correctness of the taximeter, the taximeter shall provide means to secure the connection of the taximeter to the taxi in which it is installed.
For the purpose of testing after installation, the taximeter shall be equipped with the possibility to test separately the accuracy of time and distance measurement and the accuracy of the calculation.
A taximeter and its installation instructions specified by the manufacturer shall be such that, if installed according to the manufacturer’s instructions, fraudulent alterations of the measurement signal representing the distance travelled are sufficiently excluded.
The general essential requirement dealing with fraudulent use shall be fulfilled in such a way that the interests of the customer, the driver, the driver’s employer and the fiscal authorities are protected.
A taximeter shall be designed so that it can respect the MPEs without adjustment during a period of one year of normal use.
The taximeter shall be equipped with a real-time clock by means of which the time of the day and the date are kept, one or both can be used for automatic change of tariffs. The requirements for the real-time clock are:
— The timekeeping shall have an accuracy of 0.02%;
— The correction possibility of the clock shall be not more than 2 minutes per week. Correction for summer and wintertime shall be performed automatically;
— Correction, automatic or manually, during a trip shall be prevented.
The values of distance travelled and time elapsed, when displayed or printed in accordance with this Directive, shall use the following units:
Distance travelled:
— in the United Kingdom and Ireland: until the date which will be fixed by these Member States according to Article (1)(b) of Directive 80/181/EEC: kilometres or miles;
— in all other Member States: kilometres.
Time elapsed:
— seconds, minutes or hours, as may be suitable; keeping in mind the necessary resolution and the need to prevent misunderstandings.
CONFORMITY ASSESSMENT
The conformity assessment procedures referred to in Regulation 7 that the manufacturer can choose between are:
B followed by F, or
B followed by D, or
H1.
SCHEDULE MI-008A MATERIAL MEASURES of LENGTH
The relevant essential requirements of Schedule III, the specific requirements of this Schedule and the conformity assessment procedures listed in this Schedule, apply to material measures of length defined below. However, the requirement for the supply of a copy of declarations of conformity may be interpreted as applying to a batch or consignment rather than each individual instrument.
DEFINITIONS
Material measure of length
An instrument comprising scale marks whose distances are given in legal units of length.
SPECIFIC REQUIREMENTS
Reference Conditions
1 For tapes of length equal to or greater than five metres, the maximum permissible errors (MPEs) are to be met when a tractive force of fifty newtons or other force values as specified by the manufacturer and marked on the tape accordingly, or in the case of rigid or semi-rigid measures no tractive force is needed, is applied.
2 The reference temperature is 20°C unless otherwise specified by the manufacturer and marked on the measure accordingly.
MPEs
The MPE, positive or negative in mm, between two non-consecutive scale marks is (a + bL), where:
— L is the value of the length rounded up to the next whole metre; and
— a and b are given in Table 1 below.
When a terminal interval is bounded by a surface, the MPE for any distance beginning at this point is increased by the value c given in Table 1.
| Accuracy Class | a (mm) | b | c (mm) |
| I | 0.1 | 0.1 | 0.1 |
| II | 0.3 | 0.2 | 0.2 |
| III | 0.6 | 0.4 | 0.3 |
| D — special class for dipping tapes^(1) Up to and including 30 m ^(2) | 1.5 | zero | zero |
| S — special class for tank strapping tapes For each 30 m length when the tape is supported on a flat surface | 1.5 | zero | zero |
- (1) Applies to the tape/dip weight combinations.
- (2) If the nominal tape length exceeds 30 m, an additional mpe of 0.75 mm shall be permitted for each 30 m of tape length.
Table 1
Dip tapes may also be of Classes I or II in which case for any length between two scale marks, one of which is on the sinker and the other on the tape, the mpe is ± 0.6 mm when application of the formula gives a value of less than 0.6 mm.
The MPE for the length between consecutive scale marks, and the maximum permissible difference between two consecutive intervals, are given in Table 2 below.
| Length i of the interval | MPE or difference in millimetres according to accuracy class | MPE or difference in millimetres according to accuracy class | MPE or difference in millimetres according to accuracy class |
| I | II | III | |
| i ≤ 1 mm | 0. 1 | 0.2 | 0.3 |
| 1 mm < i ≤ 1 cm | 0.2 | 0.4 | 0.6 |
Table 2
Where a rule is of the folding type, the jointing shall be such as not to cause any errors, supplementary to those above, exceeding: 0.3 mm for Class II, and 0.5 mm for Class III.
Materials
1 Materials used for material measures shall be such that length variations due to temperature excursions up to ± 8°C about the reference temperature do not exceed the MPE. This does not apply to Class S and Class D measures where the manufacturer intends that thermal expansion corrections shall be applied to observed readings where necessary.
- Measures made from material whose dimensions may alter materially when subjected to a wide range of relative humidity, may only be included in Classes II or III.
Markings
The nominal value shall be marked on the measure. Millimetre scales shall be numbered every centimetre and measures with a scale interval greater than 2 cm shall have all scale marks numbered.
CONFORMITY ASSESSMENT
The conformity assessment procedures referred to in Regulation 7 that the manufacturer can choose between are:
F1, or
D1, or
B followed by D, or
H, or
G.
SCHEDULE MI-008B CAPACITY SERVING MEASURES
The relevant essential requirements of Schedule III, the specific requirements of this Schedule and the conformity assessment procedures listed in this Schedule, apply to capacity serving measures defined below. However, the requirement for the supply of a copy of declarations of conformity may be interpreted as applying to a batch or consignment rather than each individual instrument. Also, the requirement for the instrument to bear information in respect of its accuracy shall not apply.
DEFINITIONS
Capacity serving measure
A capacity measure (such as a drinking glass, jug or thimble measure) designed to determine a specified volume of a liquid (other than a pharmaceutical product) which is sold for immediate consumption.
Line measure
A capacity serving measure marked with a line to indicate nominal capacity.
Brim measure
A capacity serving measure for which the internal volume is equal to the nominal capacity.
Transfer measure
A capacity serving measure from which it is intended that the liquid be decanted prior to consumption.
Capacity
The capacity is the internal volume for brim measures or internal volume to a filling mark for line measures.
SPECIFIC REQUIREMENTS
Reference Conditions
- Temperature: the reference temperature for measurement of capacity is 20°C.
This document does not substitute the official text published in the Irish Statute Book. We accept no responsibility for any inaccuracies arising from the transcription of the original into this format.